

Showing 1 - 1 out of 1
Page 1 out of 1
Product Name | Quantity | Total Price | Action |
---|---|---|---|
UTREL30-16 Kit 1
|
|
$199.00 |
Features
Includes the UTREL30-16 logger with the ST10M-15 external probe, buffer bottle, wall mount bracket and NIST calibration certificate.
The LogTag® UTREL30-16 Dual-Channel Low Temperature Recorder measures and stores up to a total 32,000 temperature readings.
It features a large easy-to-read display and is designed to show ‘at a glance’ if temperature excursions have occurred during the current day and up to the previous 29 days.
The alarm is triggered if temperature readings are outside pre-set alarm limits which is indicated by a large X on the display and, if configured, the audible alarm will sound.
(Silica sand recommended for temperatures below -40ºC for Buffer Assembly)
ST10M-15 external probe is used with Buffer bottle
- Kit includes: UTREL30-16, (up to 2) ST10M-15 External Probe with (up to 2) buffer bottle, Wall Mount Bracket & NIST Calibration Certificate*
- Reset Min/Max temperature reading at any time
- Audible alarm & red LED signals a temperature alarm
- View alarm status for up to 30 days at a glance
Weight: 0.25 lbs
Product Model | UTREL30-16 |
External Sensor Measurement Range | -90°C to +40°C (-130°F to +104°F) |
Operating Temperature Range | -30°C to +60°C (-22°F to +140°F) 0-99% RH (non-condensing) |
Storage Temperature Range | 0°C to +40°C (32°F to +104°F) |
Rated Temperature Resolution | Better than 0.1°C (0.2°F) for measurements -80°C to 10°C (-112°F to 50°F) Better than 0.2°C (0.4°F) for measurements 10°C to 20°C (50°F to 68°F) and -90°C to -80°C (-130°F to -112°F) Better than 0.3°C (0.5°F) for measurements 20°C to 30°C (68°F to 86°F) Better than 0.5°C (0.9°F) for measurements 30°C to 40°C (86°F to 104°F) LogTag Analyzer currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®. |
Sensor Reaction Time | Typically less than 2 minutes(T90) in moving air (1m/s) for ST10S type |
Rated Temperature Reading Accuracy | Better than ±0.5°C for -20°C to +10°C (-4°F to 50°F) Better than ±0.7°C for -50°C to -20°C (-58°F to -4°F) and +10°C to +30°C (50°F to 86°F) Better than ±1°C for -80°C to -50°C (-112°F to -58°F ) and +30°C to +40°C (86°F to 104°F) Better than ±1.8°C for -90°C to -80°C (-130°F to -112°F) |
Probe Connectivity | Up to 2 |
Probe Compatibility | All ST10 precision thermistor external probes can be interchangeably connected to the UTREL30-16 Please refer to the Accessories section for compatible External Probes. Different probe models can be used for each channel. Not compatible with ST100 sensors |
Recording Capacity | 32,000 total log memory. Provides 16,000 temperature readings per channel Day summary statistics memory (for display on LCD): up to 30 days of Min/Max/duration and alarm trigger statistics |
Sampling Interval | Configurable from 30 seconds to 18 hours. (Typically 5 minute logging) |
Alarms | Up to 6 configurable alarms for each channel Possible to set alarm separately for Channel 1 and Channel 2 |
Logging Start Options | Push button start (with configurable start delay from 1 minute to 72 hours) or specific date & time |
Download Time | Typically with full memory (32,000 readings) in less than 20 seconds |
Environmental | IP61 (when vertically mounted or hung with USB dust-cap in place) |
Power Source | 5V via micro USB socket or 2 x 1.5V AAA batteries. CR2032 LiMnO2 coin cell (backup logging function only) |
Battery Life | AAA battery up to 12 months – if unit is operated without power supply |
Real Time Clock | Built-in real time clock Typical accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day) Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C) |
Connection Interface | USB 2.0 / Micro USB Socket |
Standard Remote Probe Cable Lengths | Standard: 1.5m (4’11”) Extended: 3m (9’10”) (recommended maximum) |
Remote Probe Cable Type | PTFE (FDA food contact rated) coaxial |
Size | 141mm(H) x 72mm(W) x 18mm(T) |
Weight | 156g with AAA batteries, (132g – excluding probe and AAA batteries) |
Case Material | Polycarbonate |